1
/
of
5
Hydrogen Peroxide Low-temperature Plasma Sterilizer--SD-CPS(cassette type) 120L
Hydrogen Peroxide Low-temperature Plasma Sterilizer--SD-CPS(cassette type) 120L
Brief introduction of principle:
Hydrogen peroxide low temperature plasma sterilization system is hydrogen peroxide (H₂O₂),
in low temperature environment through the electromagnetic field of hydrogen peroxide (H₂O₂) to form hydrogen peroxide (H₂O₂) plasma, combined with hydrogen peroxide (H₂O₂) gas and hydrogen peroxide (H₂O₂) plasma for indoor items at low temperature, dry sterilization, and can effectively dissociate residual hydrogen peroxide (H₂O₂).
Technical Data
Technical Data
NO. | Item | Description of technical parameters | ||
1 | Model and weight | SD-CPS120 | ||
About 288 KG (net weight) | ||||
2 | Cavity structure and material |
The cavity structure is rectangular, the cavity material is made of 304 stainless steel, 3mm thickness, square structure, longer service life. The glow electric ion grid is aluminum alloy 5052 with a thickness of 3mm. |
||
3 | Inner material |
hexahedral PTC silicone plate is heated, fully independent temperature control, good thermal conductivity, uniform temperature, the inner tank is equipped with sterilization temperature double protection, effectively guarantee the sterilization effect and the safety of equipment and personnel. |
||
4 | The sterilization room door |
Vertical automatic drive lifting door is adopted, and the door is equipped with anti-clamp hand design safety system to ensure the personal safety of equipment and medical personnel. |
||
The new cassette plasma (right picture) has a foot-sensing open and close the door function |
||||
5 | sterilizing agent |
A cassette packed in 58% hydrogen peroxide had 12 capsules. The capacity of the cartridge less than 3ml can be used at a time |
||
6 | The method of sterilization |
Hydrogen peroxide plasma technology: Hydrogen peroxide online purification technology purithe concentration of hydrogen peroxide to more than 93%, with no residual hydrogen peroxide and no impact on the environment. |
||
7 | The sterilization speed |
Rapid sterilization time 25min; standard sterilization time 35min and enhanced sterilization time 55min |
||
8 | The sterilization temperature | 50℃±5℃ | ||
9 | screen size | 7 inch high resolution color touch screen | ||
Master controller PLC |
New program design, display working process parameters, convenient to operate, easy to understand, fully realize the control accuracy, standardization, safety, reliability; |
|||
10 | Micro Printer |
Print multiple records, record sterilization parameters, physical curve of sterilization process, record clearly, long service life |
||
The sterilization records are kept |
Computer printing provides sterilization time, cycle mode, display pressure, sterilization process, merger and sterilization physical graph printing function. The original data can be saved by large capacity storage, with a complete sterilization record of no less than 8000 times, permanent equipment alarm record, printed record can be saved for more than five years, with the function of power off memory. |
|||
11 | Pressure transmitter | Good accuracy, high reliability, and stable output | ||
12 | Temperature sensor | Dachang brand, good stability, small measurement error | ||
13 | sterilizing filter | Ultra-fine sterile filtration, the sterilization rate reached 100% | ||
14 | RF power source | Frequency 50 KHz, maximum power 500W | ||
15 | electromagnetic valve | High corrosion resistant material, reliable action life | ||
16 |
Chamber temperature heating power |
1800w, no preheat is required | ||
17 | vacuum pump |
The rotary plate vacuum pump with extremely high vacuum degree and H202 corrosion resistance is adopted, and the evacuation rate reaches 16 cubic meters / hour, and the |
||
operating sound is less than 65dB | ||||
18 |
Vacuum pump phase sequence protector |
A high vacuum phase sequence protector is provided to prevent the change, resulting in the reversal of vacuum pump and causing equipment failure |
||
19 | Oil mist filter |
The product has an exhaust oil mist filtration system, which can recover the oil mist, avoid the oil mist into the air, and through the pump suction, the pump oil back to the pump reuse to reduce fuel consumption. |
||
20 | alarm system |
System automatic detection function , real -time monitoring of the sterilization process , if any abnormal , the system automatic alarm , and display the cause of failure in Chinese ; |
||
21 |
electro-heating standing-temperature cultivator |
Temperature: 56℃ Culture time: 48 hours | ||
test method | Biometritoring ofFatobacillus thermophila | |||
22 |
The generation of the plasma species |
Single AC discharge mode, to ensure 100% plasma discharge success |
||
23 | The sterilization residue | Non-toxic damage, no residue | ||
24 | Impact on the environment | Safety has no impact | ||
25 | Impact on the medical staff |
The ultraviolet light caused by the plasma is completely shielded, completely isolated from the operator, and is invisible to the naked eye. |
||
26 | Safety ofthe sterilant |
Hydrogen peroxide is ionizes by radiofrequency glow to produce water and oxygen with no residue. |
||
27 |
The consumables are fully equipped |
Plasma special indicator card, adhesive tape, packaging bag, sterilizer, biological indicator, biological incubator, etc. shall be uniformly supplied by the manufacturer. |
||
28 |
Cavity size (allowable deviation ± 5%) |
710*480*380mm; | ||
29 |
Outer box size (allowable deviation ± 5%) |
860*790*1600mm; | ||
30 | Power supply and power | 380v | ||
3.8KVA | ||||
31 | scope of application |
Suitable for resistant to vacuum negative pressure and hydrogen peroxide gas and high temperature sensitive devices, not suitable for powder, oil and cotton devices. |
||
32 | The sterilization effect |
2m Teflon chamber with lmm inner diameter and lmm stainless steel chamber with 500mm. Meet the sterility requirements after sterilization. |
Function introduction and detailed presentation video:
Sterilizer Operating Instructions
CERTIFICATES:


SADA MEDICAL Workshop:

Catalog, Brothures and User Manual Please refer below Link:
https://www.sadamedical.com/pages/catalog-download
Sada Medical is a professional supplier of steam autoclaves, low temperature plasma autoclaves, ethylene oxide sterilizers, dry ovens, consumables, we have exported to all over the world with affordable prices.
If you're interested in our products, please feel free to contact us by Below Contact Form, looking foward to hearing from you soon.
For more product information, please click the button below to download and view!





[storeify-formbuilder-3217]