Skip to product information
1 of 3

Batch Control Helix Test 4 Bars with Chemical Indicator Holder

Batch Control Helix Test 4 Bars with Chemical Indicator Holder

FEATURES: 

Batch Control Helix Test is suitable for every sterilization process in a loaded autoclave. It can also be used to ensure maximum steam penetration where hollow instruments, or instruments with small lumina, are concerned.

*Size: 1.5 Meters
*Material: ABS heat resistant plastic, chemical heat sensitivity agent
*Operation conditions: 134 C/2 min, 121 C/10 min

FAILURE CONDITIONS:

*Sterilization temperature too low, sterilization holding time too short.

*In sufficient vacuum in depth and in number of vacuum pulses, insufficient air removal, insufficient steam penetration, leakage of piping/valves/door seals.

*Detection of presence of small volume inert gases in steam supply, detection of excessive amounts of condensate