Hydrogen Peroxide Low-temperature Plasma Sterilizer--SD-CPS(cassette type) 120L

Hydrogen Peroxide Low-temperature Plasma Sterilizer--SD-CPS(cassette type) 120L

Brief introduction of principle:

Hydrogen peroxide low temperature plasma sterilization system is hydrogen peroxide (H₂O₂),
in low temperature environment through the electromagnetic field of hydrogen peroxide (H₂O₂) to form hydrogen peroxide (H₂O₂) plasma, combined with hydrogen peroxide (H₂O₂) gas and hydrogen peroxide (H₂O₂) plasma for indoor items at low temperature, dry sterilization, and can effectively dissociate residual hydrogen peroxide (H₂O₂).

Technical Data

NO. Item Description of technical parameters
1  Model and weight SD-CPS120
About 288 KG (net weight)
2  Cavity structure and material The cavity structure is rectangular, the cavity material is made of
304   stainless   steel,   3mm  thickness,   square   structure,  longer
service life. The glow electric ion grid is aluminum alloy 5052
with a thickness of 3mm.
3  Inner material hexahedral   PTC   silicone   plate   is   heated,   fully   independent
temperature    control,    good    thermal    conductivity,    uniform
temperature,   the   inner   tank   is   equipped   with   sterilization
temperature    double    protection,    effectively    guarantee    the
sterilization effect and the safety of equipment and personnel.
4  The sterilization room door Vertical automatic drive lifting door is adopted, and the door is
equipped with anti-clamp hand design safety system to ensure the
personal safety of equipment and medical personnel.
The new cassette plasma (right picture) has a foot-sensing open
and close the door function
5  sterilizing agent A  cassette packed  in  58% hydrogen peroxide had  12  capsules.
The capacity of the cartridge less than 3ml can be used at a time
6  The method of sterilization Hydrogen peroxide plasma technology: Hydrogen peroxide online
purification    technology   purithe    concentration    of   hydrogen
peroxide to more than 93%, with no residual hydrogen peroxide
and no impact on the environment.
7  The sterilization speed Rapid sterilization time 25min; standard sterilization time 35min
and enhanced sterilization time 55min
8  The sterilization temperature 50℃±5℃
9  screen size 7 inch high resolution color touch screen
Master controller PLC New program design, display working process parameters,
convenient to operate, easy to understand, fully realize the control
accuracy, standardization, safety, reliability;
10  Micro Printer Print multiple records, record sterilization parameters, physical
curve of sterilization process, record clearly, long service life
The sterilization records are
kept
Computer printing provides sterilization time, cycle mode,
display pressure, sterilization process, merger and
sterilization physical graph printing function. The original
data can be saved by large capacity storage, with a complete
sterilization record of no less than 8000 times, permanent
equipment alarm record, printed record can be saved for
more than five years, with the function of power off memory.
11  Pressure transmitter Good accuracy, high reliability, and stable output
12  Temperature sensor Dachang brand, good stability, small measurement error
13  sterilizing filter Ultra-fine sterile filtration, the sterilization rate reached 100%
14  RF power source Frequency 50 KHz, maximum power 500W
15  electromagnetic valve High corrosion resistant material, reliable action life
16  Chamber temperature heating
power
1800w, no preheat is required
17  vacuum pump The rotary plate vacuum pump with extremely high vacuum
degree  and  H202  corrosion  resistance  is  adopted,  and  the
evacuation  rate  reaches  16  cubic  meters  /  hour,  and  the
operating sound is less than 65dB
18  Vacuum pump phase
sequence protector
A high vacuum phase sequence protector is provided to prevent
the change, resulting in the reversal of vacuum pump and causing
equipment failure
19  Oil mist filter The product has an exhaust oil mist filtration system, which can
recover the oil mist, avoid the oil mist into the air, and through the
pump suction, the pump oil back to the pump reuse to reduce fuel
                                         consumption.
20  alarm system         System  automatic  detection  function , real -time
         monitoring  of the  sterilization process ,  if any
abnormal , the  system automatic alarm , and display the
                         cause  of failure  in  Chinese ;
21  electro-heating

standing-temperature

cultivator
Temperature: 56℃ Culture time: 48 hours
test method Biometritoring ofFatobacillus thermophila
22  The generation of the plasma
species
Single AC discharge mode, to ensure 100% plasma discharge
                                         success
23  The sterilization residue Non-toxic damage, no residue
24  Impact on the environment Safety has no impact
25  Impact on the medical staff The ultraviolet light caused by the plasma is completely shielded,
completely  isolated  from  the  operator,  and  is  invisible  to  the
naked eye.
26  Safety ofthe sterilant Hydrogen peroxide is ionizes by radiofrequency glow to produce
                         water and oxygen with no residue.
27  The consumables are fully
equipped
Plasma  special  indicator  card,  adhesive  tape,  packaging  bag,
sterilizer, biological indicator, biological incubator, etc. shall be
uniformly supplied by the manufacturer.
28  Cavity size (allowable
deviation ± 5%)
710*480*380mm
29  Outer box size (allowable
deviation ± 5%)
860*790*1600mm
30  Power supply and power 380v
3.8KVA
31  scope of application Suitable for resistant to vacuum negative pressure and hydrogen
peroxide gas and high temperature sensitive devices, not suitable
for powder, oil and cotton devices.
32  The sterilization effect 2m Teflon chamber with lmm inner diameter and lmm stainless
steel chamber with 500mm. Meet the sterility requirements after
sterilization.

Function introduction and detailed presentation video:

Sterilizer Operating Instructions

CERTIFICATES:

 

SADA MEDICAL Workshop:


Catalog, Brothures and User Manual Please refer below Link:

https://www.sadamedical.com/pages/catalog-download

Sada Medical is a professional supplier of steam autoclaves, low temperature plasma autoclaves, ethylene oxide sterilizers, dry ovens, consumables, we have exported to all over the world with affordable prices.

If you're interested in our products, please feel free to contact us by Below Contact Form, looking foward to hearing from you soon.

For more product information, please click the button below to download and view!

 

Просмотреть всю информацию
[storeify-formbuilder-3217]